FCT-01 Catheter Flow Rate and Pressure Tester During Power Injection adopts the principle of high-pressure hydraulic test, which applies pressure to the test sample through constant water pressure, and maintains it for a certain period of time to test the flow and pressure of the sample under this condition. FCT-01 is suitable to the determination of flow rate and pressure of medical sterile and single-use catheters.
FCT-01 Catheter Flow Rate and Pressure Tester During Power Injection adopts the principle of high-pressure hydraulic test, which applies pressure to the test sample through constant water pressure, and maintains it for a certain period of time to test the flow and pressure of the sample under this condition. FCT-01 is suitable to the determination of flow rate and pressure of medical sterile and single-use catheters.
FCT-01 Catheter Flow Rate and Pressure Tester During Power Injection
Product Features
-7” colourful touch screen with intelligent data processing function
-With user-defined test functions, data analysis and reporting functions
-Be calibrated automatically
-Built-in high-pressure solenoid valve, high-pressure stainless steel pipeline, to ensure high-pressure test environment
-AD acquisition speed is 500KHz, internal code is 500,000; high-precision clock chip, power-off storage
-Microcomputer control dynamic display; overload alarm shutdown protection
-Special air pressure to hydraulic piping is used to ensure no leakage of each part
-Support various specifications specimen
-Automatic data storage and power-down memory function
-With a powerful data storage function, which can store up to 50,000 test records
Technical Specifications
Specifications | FCT-01 |
Test Range | 0~10Mpa |
Accuracy | <2% |
Flow Rate | 0~1000ml |
Water Bath Temperature Control | 15~55℃ |
Gas Supply Pressure | 0.4~0.7MPa (Not in supply scope) |
Port of Gas Supply | Φ8 mm PU Tubing |
Power Supply | 220V, 50Hz |
Net Weight | 70Kg |
Applications
Basic Application | Flow rate and pressure tests during power injection of central venous catheters |
Flow rate and pressure tests during power injection of balloon dilatation catheters | |
Flow rate and pressure tests during power injection of over-needle peripheral catheters |
Standards
ISO10555-1, YY0285.1
Configuration
Standard Configuration: Instrument, Booster pump
Note: 1.The gas supply port of the instrument is Φ8 mm PU Tubing;
2.Customers will need to prepare for gas supply.