LSST-04 Catheter Burst Pressure tester, adopting positive pressure test principle, is applicable to determination of rated burst pressure (RBP), balloon fatigue for freedom from leakage and damage on inflation, balloon deflation time, the relationship between the balloon diameter and the inflation pressure.
LSST-04 Catheter Burst Pressure tester, adopting positive pressure test principle, is applicable to determination of rated burst pressure (RBP), balloon fatigue for freedom from leakage and damage on inflation, balloon deflation time, the relationship between the balloon diameter and the inflation pressure.
Catheter Burst Pressure tester
Product Features
-Embedded high-speed microcomputer chip control
-7" colourful touch screen, display test data in real time
-Various test modes available
-Equipped with high-pressure booster system, test pressure up to 10MPA
-With high voltage protection device to ensure safety
-Pressure stabilization and pressure maintaining system to meet standard test requirements
-Fluid-filled test chamber, providing a 37°C temperature-controlled environment
-Luer taper for testing various types of catheters
-Over-voltage safety protection, power-down memory and other safety designs to ensure the stability
-Equipped with standard RS232 interface, which is convenient to PC connection and data transmission
-With a powerful data storage function, which can store up to 50,000 test records
-The test results judged automatically, test data can be printed
Technical Specifications
Specifications | LSST-04 |
Test Range | 0~10Mpa |
Accuracy | 0.1% FS |
Connector | Luer taper |
Gas Supply Pressure | 0.7~0.8MPa (Not in supply scope) |
Port of Gas Supply | Φ6 mm PU Tubing |
Power Supply | 220V, 50Hz |
Net Weight | 30Kg |
Applications
Basic Application | Test for burst pressure under static conditions | Sterile and single-use catheters |
Test for rated burst pressure (RBP) | Balloon dilatation catheters | |
Balloon fatigue test for freedom from leakage and damage on inflation | ||
Test for balloon deflation time | ||
Test for balloon diameter to inflation pressure |
Standards
ISO10555-1, ISO10555-4, YY0285.1, YY0285.4
Configuration
Standard Configuration: Instrument, Booster pump, Micro printer
Note: 1.The gas supply port of the instrument is Φ8 mm PU Tubing;
2.Customers will need to prepare for gas supply.